Analysis of the returned generator was performed and burn marks were observed on the pulse generator case, indicating that the pulse generator may have been exposed to an electro-cautery tool during device implant, which may have been a contributing factor.A reset of the pulsedisable bit in the generator memory was performed to allow for an output to once again be provided by the generator for subsequent testing.The battery of the returned generator was measured as 3.048 volts showing an ifi=no condition.The downloaded data revealed that 0.938% of the battery had been consumed.Other than the observed pulse disablement and burn marks, there were no performance or any other type of adverse conditions found with the pulse generator.The cause for the reported premature battery depletion was due to electromagnetic induction (emi) or electrostatic discharge (esd) transients.Premature battery depletion can result from the use of electrocautery during surgery or an electrostatic discharge imparted to the generator through the torque wrench used.
|