(b)(4).Method: the complaint rt380 adult dual-heated evaqua2 breathing circuit was returned to fisher & paykel healthcare (fph) in (b)(4) for investigation.The inspiratory and expiratory heater wire of the circuit was resistance tested using a calibrated multimeter.Results: the resistance test revealed that the inspiratory and expiratory heater wire on the returned circuit was within specification.A lot check revealed no other complaints of this nature for lot 150304.Conclusion: following this incident, an fph representative visited the hospital and obtained further information and offered advice to address the reported condensation issue.The fph representative noticed that the temperature in the affected room was 25 to 27 degrees celsius, which is very warm.This temperature is at the upper end of the mr850 humidifier operating conditions of 18 to 26 degrees celsius, stated in the mr850 user instruction.A high room temperature potentially can contribute to excessive condensation, thus room temperature monitoring is essential.Condensate in the humidification system, is an expected side effect of heated pass-over humidification systems in many conditions, and may vary between light misting to water droplets that form on the wall of cool breathing circuit tubing.The amount of condensate in the ventilation system is influenced by a number of multiple setup and environmental factors.Since the fph representative visited the hospital, we have received no further reports relating to condensation issues.The user instructions that accompany the mr850 state the following: operating conditions: recommended ambient temperature 18 - 26 °c.Caution: if operating in ambient temperatures outside the recommended range, consult your local fisher & paykel healthcare representative or consult the technical manual.The user instructions that accompany the rt380 state the following: set appropriate ventilator alarms.Check breathing circuits for condensation every 6 hours and drain if required.
|