Model Number 2090AA
Device Problem Programming Issue (3014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/05/2015
Event Type  malfunction  
Event Description
It was reported that the programmer would not start.The programmer was returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: analysis confirmed the customer comment that the programmer would not start, it did not proceed to a functioning screen which would allow interrogation.A reconfiguration was completed and the software reloaded to resolve.It was further noted that the stylus tip was separating from the main pen body and that the system fan was intermittently noisy.Both the stylus and the fan were replaced.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC INC.
8200 coral sea st ne
mounds view,mn 5511 2
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4829102
MDR Text Key5838821
Report Number2182208-2015-01514
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090AA
Device Catalogue Number2090AA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received03/05/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
2067 RADIOFREQUENCY PROGRAMMER HEAD
-
-