• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE MICROFX OCD UNIVERSAL DRILL REAMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE MICROFX OCD UNIVERSAL DRILL REAMER Back to Search Results
Catalog Number 234-200-200
Device Problems Break; Tip
Event Date 06/02/2015
Event Type  Malfunction  
Event Description

It was reported that the tip broke off inside the patient during surgery; however, it was retrieved.

 
Manufacturer Narrative

Additional information will be provided once the investigation has been completed.

 
Manufacturer Narrative

The reported device was not received for investigation; therefore the reported failure cannot be confirmed. The complaint will be closed without a detailed investigation report and based on probable root cause. In the event that the device is received, the complaint will be reopened and the investigation will be updated with the new results. Probable root cause for the reported failure involving this device could have been caused by: design: drill too dull or dulls quickly, inefficient flute design, requires excessive force to advance. Drill not in alignment with guide. Drill allowed to extend too far out of guide. Process: drill manufactured out of specification drill guide manufactured out of specification application: excessive axial force on drill. Run drill in reverse. Reuse/resterilization of single-use device, or use of a single drill on multiple defects. Guide is translated and/or rotated during drilling; drill not in alignment with guide. User does not use snap cap, drills too deep. Drill is started/stopped while in bone. The product was not returned for investigation therefore the reported failure mode was not confirmed. The failure mode will be monitored for future reoccurrence.

 
Event Description

It was reported that the tip broke off inside the patient during surgery; however, it was retrieved.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMICROFX OCD UNIVERSAL DRILL
Type of DeviceREAMER
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose , CA 95138
4087542000
MDR Report Key4867061
Report Number0002936485-2015-00526
Device Sequence Number1
Product CodeHTO
Report Source Manufacturer
Source Type Other,OTHER
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/02/2015
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received06/24/2015
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue Number234-200-200
Device LOT Number56877
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/02/2015
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-