• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number R5C8320
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/29/2015
Event Type  malfunction  
Event Description
It was reported that a patient experienced an electric shock during set-up of a homechoice device.The patient did not sustain any burn or other injury.The patient ended therapy and the call center advised to use manual supplies to complete therapy.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).Additional information: the device was returned and an evaluation was complete.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Full functional testing, electrical safety testing, calibration and simulated therapy were performed with no issues noted.The reported condition was not verified.The power supply fuse and power cord were replaced preventively.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4896420
MDR Text Key6535258
Report Number1416980-2015-28005
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeTU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative,company representati
Reporter Occupation Other
Type of Report Initial
Report Date 06/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberR5C8320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received08/05/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age34 YR
Patient Weight53
-
-