It was reported that the vns patient underwent generator and lead replacement surgery on (b)(6) 2015.The generator was near end of service but was unable to be interrogated at the time of the procedure believed to be due to end of service.The lead was replaced prophylactically.During the procedure, the surgeon replaced the patient's generator and subsequently used electrocautery to replace the lead.When the device was tested after the device replacements, the replacement generator showed an end of service condition.The device was tested prior to the use of electrocautery and the device showed battery status as ok.Another generator was used to complete the procedure.The explanted devices and the opened but unused generator were returned to the manufacturer for analysis.The end of service condition and failure to program with the explanted generator was the result of normal, expected battery depletion.The device performed according to functional specifications.Analysis concluded that no abnormal performance or any other type of adverse condition was found.Review of the data from the opened but unused generator indicated that the pulse disabled byte was set to a value that represents a vbat
|