Catalog Number 0848000000 |
Device Problem
Smoking (1585)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/17/2015 |
Event Type
malfunction
|
Event Description
|
It was that power switch in the cast cutter was smoking during a cast removal procedure at the user facility.The procedure was completed successfully.No delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
Failure analysis is in progress; additional information will be submitted once the quality investigation is completed.
|
|
Manufacturer Narrative
|
The reported event, unit started smoking, was not duplicated.However, it was confirmed as a non-stryker switch and cord assembly were found installed in the device, which can cause or contribute to the device smoking.Through inspection, the service technician also noted that the device failed to run.
|
|
Event Description
|
It was that power switch in the cast cutter was smoking during a cast removal procedure at the user facility.The procedure was completed successfully.No delay, no medical intervention and no adverse consequences were reported with this event.
|
|