Model Number 2090AA
Device Problem Device Maintenance Issue (1379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/16/2015
Event Type  malfunction  
Manufacturer Narrative
This report is based solely on device analysis.If additional relevant information is received, a supplemental report will be submitted.Product event summary: analysis confirmed the reported event; overlay was broken.Analysis also found the ecg (electrocardiogram) connector was loose.The keyboard hinge was broken and the latch tabs were broken off of the power cord door.Concomitant product(s): product id: 229047 analyzer.(b)(4).
 
Event Description
It was reported the screen needs to be repaired.The programmer was returned to the manufacturer for repair and calibration, analyzed and tested out of specification.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4966623
MDR Text Key23391114
Report Number2182208-2015-02333
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Report Date 05/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090AA
Device Catalogue Number2090AA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received05/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-