A health care provider reported via a manufacturing representative that high impedances were measured during intraoperative testing.The extensions were removed from the implantable neurostimulator (ins), placed in opposite ins channels, and then placed back in their original channels.After moving the extensions around, the impedance issue seemed to be resolved.A week after implant, impedance of contact nine was measured to be above 40,000 ohms.A revision was done and the extensions were disconnected from the ins and switched around.Impedances were then measured to be normal.The extensions were then switched to their original position and impedance was measured to be high on contact nine again.The manufacturing representative had expected contact one to have high impedance following the swap.The extensions were removed again, cleaned and reinserted into the ins.Impedances were then measured to be normal.The issue was resolved at the time of this report.
|