Model Number 15880
Device Problem Kinked (1339)
Patient Problem Hyperglycemia (1905)
Event Date 07/19/2015
Event Type  Malfunction  
Manufacturer Narrative

The product was not returned for evaluation. We are unable to confirm the reported kinked cannula or to determine whether it contributed to the reported hyperglycemia. Lot release records were reviewed and the product lot met all acceptance criteria. The omnipod¿s user guide warns "test results greater than 13. 9 mmol/l [250 mg/dl] mean high blood glucose (hyperglycemia). If you get results above 13. 9 mmol/l [250 mg/dl], but do not have symptoms of hyperglycemia, repeat the test. If you have symptoms or continue to get results above 13. 9 mmol/l [250 mg/dl], follow the treatment advice of your healthcare provider," and "if your reading is above 27. 8 mmol/l [500 mg/dl], the pdm displays ¿high check for ketones!¿ this indicates severe hyperglycemia (high blood glucose). ".

 
Event Description

The customer reported his blood glucose read "high" (> 27. 8 mmol/l) (> 500 mg/dl) and that the cannula had a small kink. The pod was worn between 4 and 24 hours.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOMNIPOD INSULIN PUMP
Type of DevicePUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
600 technology park drive
suite 200
billerica MA
Manufacturer (Section G)
INSULET CORPORATION
600 technology park drive suit
billerica MA
Manufacturer Contact
david simard
600 technology park drive
suite 200
billerica, MA 
9786007000
MDR Report Key5015973
MDR Text Key23527254
Report Number3004464228-2015-00566
Device Sequence Number1
Product Code LZG
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK122953
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER,FOREIGN
Reporter Occupation
Type of Report Initial
Report Date 08/19/2015
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/19/2015
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date09/18/2016
Device MODEL Number15880
Device Catalogue NumberZXR425
Device LOT NumberL41649
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/20/2015
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured03/15/2015
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-