• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problem Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2014
Event Type  malfunction  
Manufacturer Narrative
One power max elite motor drive unit was received on december 29, 2014 and confirmed to be serial number (b)(4).The reported complaint has been confirmed.Product failed with hand piece sensor fault and shorted out dii tester.Unit was found to have a short circuit that will cause electrical damage to the control unit.The complaint investigation has concluded the cause of the failure to be a defective electronic component for which the cause is unknown.Cause of fault and short is a defective hall board.The motor was also noted to be damaged and runs intermittently.Unit requires non warranty repair.Smith & nephew will continue to monitor for any future occurrences of this failure type.No further action is deemed necessary as a result.
 
Event Description
When the shaver was plugged in, the machine started to smoke.There were no injuries to the patient or operating room staff.It is unclear if it was shaver that caused this issue or not.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
150 minuteman road
andover, MA 01810
5123585706
MDR Report Key5017129
MDR Text Key24658830
Report Number1643264-2015-00124
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Report Date 12/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Device Lot NumberAAL58304
Other Device ID NumberAAL58304
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-