Product analysis was completed on the returned lead.During the visual analysis several areas of the outer silicone tubing appeared to be compressed and twisted.The positive and negative quadfilar coils were observed to be broken in several places.The coil breaks showed signs of wear, pitting, mechanical damage, and stress induced fracture (including rotational forces).It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting.Abrasions were noted on both the inner and outer silicone tubing.The compressed and twisted conditions observed in the quadfilar coil break areas of the lead tubing suggest the two issues may be related to one another.With the exception of the observed discontinuities and twisted/compressed lead tubing, the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Note that since a portion of the (+) white and (-) green electrode inner silicone tubes and quadfilar coils including the anchor tether and (+) white electrode were not returned, an evaluation and resulting commentary cannot be made on that portion of the lead.
|