The actual device was returned to the mfr for physical eval.External visual inspection.There were no signs of any physical damage with the received cycler.The heater tray/scale was not obstructed.Interior inspection: the right, rear panel bracket threaded fitting had become dislodged from the l-shape bracket.Two simulated treatments were performed using the received cycler, and there were no alarms that occurred during testing.A simulated treatment was performed in which the amount of fluid delivered to a pseudo-patient bag during each fill phase was weighed.The weighed fluid volumes were compared against the programmed fill value and the weighed volumes for each fill phase were determined to be within expected tolerance for the liberty cycler.An analysis of the results of two simulated treatments showed the two drain volumes were slightly out-of-specification; 3ml and 4ml respectively.The fault was corrected by tarring the load cell.After tarring the load cell, a subsequent simulated treatment was performed and the resulting drain volumes were within tolerance.The initial load cell value was within tolerance.The load cell verification was within tolerance.The valve actuation test, system air leak test, and the pressure sensor verification check passed.A device manufacturing record review was also conducted.The device history record (dhr) review showed- production floor problem report on (b)(6) 2015 (15.5.1) erratic load cell occurred rework-as per liberty cycler rework manual occurred; disconnected j16 and connected j17 sent to (b)(4) on 04/07/2015.Product labeling, material, and process controls were within specifications.
|