Model Number 14000
Device Problems Tube; Kinked
Event Date 07/25/2015
Event Type  Malfunction  
Manufacturer Narrative

The device was not returned for evaluation. We are unable to confirm the kinked cannula or to determine if it could have contributed to the reported hyperglycemia. Lot release records were reviewed and the product lot met all acceptance criteria. The omnipod user guide warns, "test results greater than 250 mg/dl mean high blood glucose (hyperglycemia). If you get results above 250 mg/dl, but do not have symptoms of hyperglycemia, repeat the test. If you have symptoms or continue to get results that fall above 250 mg/dl, follow the treatment advice of your healthcare provider. ".

 
Event Description

Customer reported high blood glucose values and a kinked cannula. She reported the following blood glucose and insulin history: (b)(6).

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOMNIPOD INSULIN PUMP
Type of DevicePUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
600 technology park drive
suite 200
billerica MA 01821
Manufacturer (Section G)
INSULET CORPORATION
600 technology park drive suit
billerica MA 01821
Manufacturer Contact
david simard
600 technology park drive
suite 200
billerica , MA 01821
9786007000
MDR Report Key5025435
Report Number3004464228-2015-00616
Device Sequence Number1
Product CodeLZG
Report Source Manufacturer
Source Type CONSUMER
Reporter Occupation
Type of Report Initial
Report Date 07/25/2015
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/24/2015
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date09/01/2016
Device MODEL Number14000
Device Catalogue NumberZXP425
Device LOT NumberL41661
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/25/2015
Was Device Evaluated By Manufacturer? No
Date Device Manufactured03/01/2015
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-