MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number PULSAR
Device Problems Improper or Incorrect Procedure or Method (2017); Device Operates Differently Than Expected (2913)
Patient Problems Head Injury (1879); Failure of Implant (1924)
Event Date 06/20/2015
Event Type  Injury  
Manufacturer Narrative
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
It was reported that the patient suffered an impact to the head that caused damage to the device.Thereafter, the patient has been re-implanted.
 
Manufacturer Narrative
Additional information: based on the received information, the implant housing was damaged due to an accident.The device has not been received for investigation.The complaint can be re-opened when the device is available for investigation.
 
Event Description
It was reported that the patient suffered an impact to the head that caused damage to the device.The patient has been re-implanted.
 
Manufacturer Narrative
The device investigation revealed mechanical damage to the stimulator housing that is typical for severe external impact to the housing.The problems described in the recipient report appear to match the damage found.This is a final report.
 
Event Description
It was reported that the recipient had an impact to the head that caused damage to the device.The recipient underwent re-implantation surgery on (b)(6) 2015.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key5027041
MDR Text Key23908658
Report Number9710014-2015-00595
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Followup,Followup
Report Date 03/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPULSAR
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/06/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age17 YR
-
-