INSULET CORPORATION OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN Back to Search Results
Model Number 15880
Device Problems Tube; Unintended Movement
Event Date 07/27/2015
Event Type  Malfunction  
Manufacturer Narrative

The product was not returned for evaluation. The user reported that the cannula had pulled out of the infusion site. A dislodged cannula could interrupt insulin delivery and may lead to hyperglycemia. The omnipod¿s user guide warns to "check often to make sure the pod and soft cannula are securely attached and in place. A loose or dislodged cannula may interrupt insulin delivery. Verify there is no wetness or scent of insulin, where as may indicate the cannula has dislodged," "because insulin pods use only rapid-acting insulin, users are at increased risk for developing hyperglycemia (high blood glucose) if insulin delivery is interrupted," and "test results greater than 13. 9 mmol/l [250 mg/dl] mean high blood glucose (hyperglycemia). If you get results above 13. 9 mmol/l [250 mg/dl], but do not have symptoms of hyperglycemia, repeat the test. If you have symptoms or continue to get results above 13. 9 mmol/l [250 mg/dl], follow the treatment advice of your healthcare provider. " no product lot number was reported therefore no qualification records were reviewed.

 
Event Description

The customer reported her blood glucose reached over 13. 8 mmol/l (248 mg/dl) and that the cannula had slipped out of the skin.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOMNIPOD INSULIN PUMP
Type of DevicePUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
600 technology park drive
suite 200
billerica MA 01821
Manufacturer (Section G)
INSULET CORPORATION
600 technology park drive suit
billerica MA
Manufacturer Contact
david simard
600 technology park drive
suite 200
billerica , MA 
9786007000
MDR Report Key5027823
Report Number3004464228-2015-00630
Device Sequence Number1
Product CodeLZG
Report Source Manufacturer
Source Type CONSUMER,FOREIGN
Reporter Occupation
Type of Report Initial
Report Date 07/27/2015
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/25/2015
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device MODEL Number15880
Device Catalogue NumberZXR425
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/27/2015
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-