Model Number 37601
Device Problems Break (1069); High impedance (1291); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 37085-40, serial# (b)(4), product type: extension.(b)(4).
 
Event Description
The health care provider (hcp) reported via the manufacturer's representative that high impedances were measured on lead contact 3.The high impedances occurred because a replacement of the system was performed at a clinic inexperienced with deep brain stimulation and the extension was incorrectly connected to the header block of the implantable neurostimulator (ins).It was noted the extension was also accidentally cut during the replacement.A revision surgery occurred.The hcp was unable to disconnect the extension from the header block.The ins and two extensions were replaced.
 
Manufacturer Narrative
Device analysis for ins (b)(4) revealed no anomaly.Device analysis for extension (b)(4) revealed the proximal end connector not completely seated in ins connector port.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
IPG MFG SWITZERLAND
route du molliau 31
tolochenaz 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key5036331
MDR Text Key24351253
Report Number9614453-2015-02156
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Followup
Report Date 08/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2016
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age00051 YR