Catalog Number 5532-G-609
Device Problem Insufficient Information (3190)
Patient Problems Loss of Range of Motion (2032); Scarring (2061)
Event Date 08/10/2015
Event Type  Injury  
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was retained by the patient and was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Due to reduced range of motion and excessive scaring the insert was removed and replaced with a new one.
 
Manufacturer Narrative
An event regarding reduced range of motion involving a triathlon insert was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as no product was returned for evaluation.Medical records received and evaluation: not performed as no medical records were received.Device history review: indicated the devices accepted into final stock from the reported lot were free from discrepancies complaint history review: there has been no other event for the lot referenced.Conclusions: the event could not be confirmed nor the root cause determined because the devices were not returned for evaluation and insufficient medical information was provided.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.
 
Event Description
Due to reduced range of motion and excessive scaring the insert was removed and replaced with a new one.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
william hanna
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5050100
MDR Text Key24888125
Report Number0002249697-2015-02902
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Followup
Report Date 08/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Catalogue Number5532-G-609
Device Lot NumberMMMMJY
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/25/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/06/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient Weight131
-
-