COOK ENDOSCOPY HERCULES 3 STAGE BALLOON ESOPHAGEAL; KNQ, DILATOR, ESOPHAGEAL Back to Search Results
Catalog Number HBD-18-19-20
Device Problem Hole In Material (1293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/15/2015
Event Type  malfunction  
Event Description
During an esophageal dilation procedure, the physician used a cook hercules 3 stage balloon esophageal.The balloon failed to inflate when in position down the endoscope.Another device of the same type was used to complete the procedure.The device was returned to cook on 8/17/2015 for evaluation.Our evaluation determined the balloon has two pinholes near the center of the balloon.
 
Manufacturer Narrative
On 9/8/2015, we became aware that the information provided in the other remarks section was not submitted on the initial emdr.Concomitant medical products: boston scientific alliance gun handle, cook dilation syringe ((b)(4)).Investigation evaluation: our evaluation of the returned device confirmed the report.The balloon was inflated with a 60 cc syringe and an inflation handle.The balloon has two pinholes near the center of the balloon.The balloon would not hold pressure or remain inflated in this condition.There are numerous kinks along the length of the catheter.No part of the device is missing.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusions: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.A possible contributing factor to balloon material failure is inadequate lubrication of the balloon with a lubricating agent.The instructions for use direct the user to apply a lubricating agent to the balloon to allow easier passage through the accessory channel.This activity will aid in endoscopic advancement and balloon preservation.The instructions for use advise the user that negative pressure is needed to maintain balloon deflation.The application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.A balloon material failure can occur if the balloon is inflated prior to advancement through the endoscope or if the balloon is inflated while partially or fully inside the accessory channel of the endoscope.The instructions for use contain the following precaution: ¿do not pre-inflate the balloon.¿ the instructions for use state, "maintain balloon deflation with negative pressure and introduce into the accessory channel of the endoscope, advancing in short increments until the dilator is completely visualized endoscopically.Monitor endoscopically until the balloon is in the desired position within the stricture." the instructions for use contain the following warning: during dilation do not inflate balloon beyond the maximum indicated inflation pressure, as this could result in overextension or bursting of the balloon.To achieve increasingly larger balloon diameters, increase pressure as indicated on the catheter tag.Another possible contributing factor is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in a failure of the balloon material.Prior to distribution, all hercules 3 stage balloon esophageal are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HERCULES 3 STAGE BALLOON ESOPHAGEAL
Type of Device
KNQ, DILATOR, ESOPHAGEAL
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key5052739
MDR Text Key24887723
Report Number1037905-2015-00380
Device Sequence Number1
Product Code KNQ
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K061937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Report Date 07/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHBD-18-19-20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/15/2015
Device Age11 MO
Event Location Hospital
Date Manufacturer Received09/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-