Model Number IRC5PO2 |
Device Problems
Device Alarm System (1012); Gas Output Problem (1266)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional information become available, a supplemental record will be filed.
|
|
Event Description
|
The dealer states the unit has low o2 at 5 lpm, was well below 85% and not alarming.
|
|
Manufacturer Narrative
|
Unit repaired by independent repair center.Irs received 10/2/15.Reason for repair is low o2 or yellow light.The key failure is a leaking pilot valve.The verification test 1 o2 purity was 87%.The underlying cause of the lower than expected o2 purity level documented in the irs is a stuck/not shifting pilot valve.Note: the unit's audible alarm is activated when the purity falls below 73%.Verification testing results provided on the irs indicate the o2 purity was at 87%.The unit audible alarm would not be triggered as they are above 73%.This is not a malfunction of the audible alarm system.
|
|
Event Description
|
The dealer states the unit has low o2 at 5 lpm, was well below 85% and not alarming.Independent repair center irs received 10/2/15.Reason for repair is low o2 or yellow light.The key failure is a leaking pilot valve.The verification test 1 o2 purity was 87%.The underlying cause of the lower than expected o2 purity level documented in the irs is a stuck/not shifting pilot valve.Note: the unit's audible alarm is activated when the purity falls below 73%.Verification testing results provided on the irs indicate the o2 purity was at 87%.The unit audible alarm would not be triggered as they are above 73%.This is not a malfunction of the audible alarm system.
|
|
Search Alerts/Recalls
|