Event Description

It was reported that the patient has left breast cancer and will be explanted. Although surgery is likely, it has not occurred to date. Good faith attempts for further information from the physician were unsuccessful.

 
Event Description

On 10/23/2015 it was reported that the patient underwent prophylactic generator replacement and the new generator was moved to the right chest wall due to the patient's left breast cancer. Pre-operative diagnostics showed results within normal limits. The explanted generator was returned for product analysis on 11/12/2015. Product analysis is still underway and has not yet been completed.

 
Event Description

Product analysis was completed on the generator on 12/10/15. In the product analysis lab, the device output signal was monitored for more than 24 hours, while the generator was placed in a simulated body temperature environment. Results showed no signs of variation in the pulse generator's output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period. The pulse generator diagnostics were as expected for the programmed parameters. In addition, a comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications. There were no performance or any other type of adverse conditions found with the pulse generator.

 
Event Description

On (b)(6) 2016 the physician reported that there is no relationship between the patient's cancer and vns. The physician declined to provide the date the cancer was first observed.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePULSE GEN MODEL 103
Type of DeviceGENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key5079025
Report Number1644487-2015-05798
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation HEALTH PROFESSIONAL
Type of Report Initial,Followup,Followup,Followup
Report Date 08/25/2015
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received09/15/2015
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator LAY USER/PATIENT
Device EXPIRATION Date02/03/2013
Device MODEL Number103
Device LOT Number201717
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer11/12/2015
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location Other
Date Manufacturer Received01/28/2016
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured03/28/2011
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 09/15/2015 Patient Sequence Number: 1
-
-