During an implant surgery, the surgeon reported that the electrodes seemed to be "straighter" and that the electrode coils didn't have the normal turns so it was more difficult to wrap around the nerve.Diagnostics performed with this lead were recorded the impedance to be 748 ohms and 800 ohms.Due to the lower impedance value, the surgeon opted to replace the lead.The explanted lead was received on (b)(6) 2015.Analysis is underway but has not been completed to date.An impedance value greater than 600 ohms and less than 5300 ohms is considered to be within normal limits.
|