STRYKER ORTHOPAEDICS-MAHWAH PRIMARY TRITANIUM HEMI CLUSTER HOLE CUP 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
Catalog Number 502-03-52D |
Device Problems
Material Separation (1562); Sticking (1597); Manufacturing, Packaging or Shipping Problem (2975); Packaging Problem (3007)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/10/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
When completed, the evaluation summary will be submitted in a supplemental report.
|
|
Event Description
|
It was reported, "inner tyvek was stuck to outer one.Inner tyvek was separated to two sheets.So, sterility was kept barely.".
|
|
Manufacturer Narrative
|
An event regarding the tyvek lids sticking together involving a primary tritanium hemi cluster hole cup 52mm was reported.The event was confirmed.Method & results: -device evaluation and results: visual evaluation of the returned device packaging confirmed that inner blister tyvek lid was stuck to the outer lid.-medical records received and evaluation: not performed as no patient information was provided for review.-device history review: a device history review indicated that all devices that were accepted into final stock conformed to specification.-complaint history review: a complaint history review indicated that there have been no similar events for the reported lot.Conclusions: the investigation concluded that tyvek lids sticking was caused by a known packaging issue.Packaging innovation is aware of this, and has issued a memo.Due to the infrequent nature of these complaints and the greater risk that would be presented by an undersized foam, these events will not be addressed through a design change.
|
|
Event Description
|
It was reported, "inner tyvek was stuck to outer one.Inner tyvek was separated to two sheets.So, sterility was kept barely.".
|
|