Product analysis was completed on the generator on (b)(4) 2015.Various electrical loads were attached to the generator and results of diagnostic tests demonstrate that accurate resistance measurements were obtained in all instances.In the product analysis lab, the device output signal was monitored for more than 24-hrs, while the generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator铠output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.Magnet activations performed during output monitoring (at a distance of one-inch, spacer block, from the generator), demonstrate the appropriate magnet output for the programmed settings.In addition, a comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications; there were no performance or any other type of adverse conditions found with the pulse generator.Product analysis was completed on the leads on (b)(4) 2015.An analysis was performed on the returned lead portions and bare and exposed conductive coils may be a contributing factor condition that could potentially contribute to the reported low impedance.During the visual analysis of the returned 389mm portion the connector boot / inner silicone tubes appeared to be torn / pulled apart in half (at the end of the connector ring) with the quadfilar coils bare, exposed and stretched in between.During the visual analysis of the returned 56mm portion the (+) white electrode inner silicone tubing appeared have been melted through and the quadfilar coil appeared to be slightly melted.Additional melted areas were observed on both of the inner silicone tubes.Based on the obvious signs of mechanical damage on the inner silicone tubes and quadfilar coil, it is possible the thermally-damaged coil was exposed to a high temperature device such as a cauterizing tool (electrosurgical unit) during the explant of this lead (damaged coil, not a break).The abraded opening found on the outer silicone and the cut ends that were made during the explanted process, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.For the observed inner tubing fluid remnants, there was no obvious path for fluid ingress other than the cut ends that were made during the explanted process and the torn ends of the connector boot / outer / inner silicone tubes at the end of the connector ring.With the exception of the abraded opening and torn / pulled apart connector boot / inner silicone tubes, the condition of the returned lead portions is consistent with those that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no discontinuities were identified.
|