• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/17/2015
Event Type  malfunction  
Event Description
It was reported that a company representative was having difficulties to use her programming system.It was reported that the handheld computer was not functioning properly due to a suspected connection issue with the serial cable.The handheld was returned to the manufacturer on (b)(6) 2015.Analysis is underway but it has not been completed to date.
 
Event Description
Analysis of the returned handheld was completed.No anomalies associated with the returned x50 serial cable were identified during the analysis.During the analysis, it was identified that the handheld would not power on using the ac adapter.The cause for the anomaly is associated with damaged leads associated with the handheld sync connector.Because of the damage, the handheld was unable to receive power from the ac adapter.The cause for the damage to the connector leads and solder connections is most likely associated with mishandling of the device as it appears the connector detents are not being compressed when removing or manipulating the serial data cable, thus placing an extensive amount of force on the pcb and attached cable receptacle.The non-responsiveness from hhd is due to a damaged connector receptacle where the serial data cable normally attaches; receptacle became partially separated from the main pcb with some broken connector pins; most likely associated with wear or excessive cable manipulation.Analysis of the returned flash card was completed and no anomalies associated with flashcard software were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5146713
MDR Text Key28520339
Report Number1644487-2015-06090
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberMODEL 250
Device Lot Number1063524
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2015
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received10/15/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-