It was reported that high impedance was observed during the patient's initial implant surgery on (b)(6) 2015.After the new generator was implanted, diagnostics performed with the new generator and old lead out of pocket showed high impedance.The surgeon confirmed that the lead was not nicked during the surgery.The nerve, was irrigated but the diagnostics shill showed high impedance.The lead pin was reinserted into the generator and diagnostics still resulted in high impedance.The lead pin was removed and generator diagnostics with the lead resistor was performed resulting in normal results.The generator was connected to the old lead and high impedance persisted.Finally, the lead was replaced by the surgeon.Additional information was received that the high impedance was observed on system diagnostic test.Normal impedance was received once the lead was replaced.The suspect lead was received on 09/29/2015 due to lead discontinuity.Analysis is underway but has not yet been completed.
|