• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. LEAD MODEL 304

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. LEAD MODEL 304 Back to Search Results
Model Number 304-20
Device Problem High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/30/2015
Event Type  malfunction  
Event Description
It was reported that high impedance was seen during system diagnostics.The physician disabled the patient's generator due to the high impedance.The manufacturing records for the lead was reviewed and the device met all specifications prior to shipment.The physician's office has ordered x-rays but they still have not been performed.They will provide a copy of them to the manufacturer once they are taken.
 
Event Description
In a (b)(6) 2016 appointment, the patient¿s generator had normal mode current turned on.However, the patient then experienced neck pain and coughing with stimulation.The normal mode was turned back off.High impedance was again noted in this appointment.On a subsequent date, chest x-rays were stated to have been performed, and the medical professional assessed that the vns system was intact.No additional pertinent information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 304
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5159287
MDR Text Key28950946
Report Number1644487-2015-06164
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/30/2018
Device Model Number304-20
Device Lot Number203000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received10/12/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/02/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age9 YR
-
-