It was reported that during a near end-of-service battery replacement surgery, a calcified tissue was identified on the upper right portion of the generator can, below the header.The surgeon suspected that the area with calcification was a result of corrosion.The calcified tissue was chipped away from the generator.Systems diagnostics at pre-surgery and during surgery were reported to be within normal limits.There was no tissue located at the header or the welded seam of the generator can.No infection was noted in the area.The product has been received for analysis, which has not been completed to-date.
|
Analysis was completed on the returned generator.Other than typical explant procedure related observations, no surface abnormalities, such as sharp edges, header delamination, open pockets, decomposition, corrosion, or voids were noted on this device.There were white deposits on the can.A scanning electron microscopy analysis of the foreign matter indicated that it was calcium and phosphorus.There was no evidence of dried body fluid or corrosion in the connector block.There was no indication from the device that an end of service condition existed.A battery life calculation resulted in 0.50 years remaining before the elective replacement indicator flag would be set.Analysis of the generator concluded that no abnormal performance or any other type of adverse condition was found.
|