• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T167
Device Problems Migration or Expulsion of Device (1395); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Pocket Erosion (2013)
Event Date 03/02/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that at routine follow up this implantable cardioverter defibrillator (icd) was observed to have migrated.The physician elected not to intervene at the time in favor of waiting until device replacement was needed due to normal battery depletion.Several months later the patient presented for device replacement and the beginning signs of pocket erosion were observed though the device had not yet ruptured through the patient's skin; infection was suspected.The device was explanted and replaced with a new device pocket created above the previous.A small insulation breach was noted on the competitor right atrial (ra) lead that was also repaired during the procedure though no signs had previously presented.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5172539
MDR Text Key29170749
Report Number2124215-2015-12061
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date05/19/2009
Device Model NumberT167
Other Device ID NumberVITALITY 2 DR EL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received08/03/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/21/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age74 YR
-
-