Product analysis was completed on the returned portion of the lead.Analysis of the returned portion of the lead determined that the report of a kinked lead was not verified.However, a large portion of the lead assembly (body), including the electrodes, was not returned for analysis; therefore, a complete evaluation could not be performed on the entire lead product.The abraded opening found on the outer silicone tubing and the cut ends that were made during the explant process most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.For the observed inner tubing fluid remnants, there was no obvious path for fluid ingress other than the cut ends that were made during the explant process.What appeared to be white deposits were observed in various locations.Chemical analysis was performed on the deposit observed on the outer silicone tubing and identified the deposit as containing silicone.The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.The setscrew marks found on the lead connector pins provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portion were performed and no discontinuities were identified.Based on the findings in the product analysis lab, there is no evidence to suggest an anomaly with the returned portion of the device.
|