Patient was reading low battery on the nk monitor.
Took 2 (b)(6) batteries in telepack, telepack would not turn on.
Went to check to see if battery was in wrong position.
Batteries were in right but noted one battery to be very hot, barely able to touch to remove with gloves on.
Removed those two batteries and placed 2 brand new batteries in telepack, locked the door and telepack turned on.
Rhythm noted on monitor.
Patient said this is a little warm but not as hot as the last one felt.
I said let me just check it, opened the telepack battery in same side was hot again.
Placed patient on hardwire to take telepack out of service.
Walked out of the room, label on battery started to curl, walked out to nurses' station sat it down.
A second later started to smell like an electrical fire, started to slightly smoke, battery wrapper opened up black portion was bulging out of the wrapper.
Noted that the negative terminal of the nk pack was bent a gap between the negative terminal and plastic cover exists-the (b)(6) batteries were tested with a non-defective telepack.
With effort the batteries could be positioned in a way in which the connector came in contact with the plastic cover and negative battery terminal, melting of the cover occured.
Battery used was (b)(6).
|