Catalog Number 5532-G-309
Device Problems Peeled/Delaminated (1454); Unstable (1667); Insufficient Information (3190)
Patient Problems Injury (2348); No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2015
Event Type  Injury  
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was retained by the hospital and was not returned to the manufacturer.Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Surgeon revised patient with mid range instability and sheered off patella.Left knee.
 
Manufacturer Narrative
An event regarding revision due to instability involving a triathlon insert was reported.The event was not confirmed.Method & results: -device evaluation and results: not performed as no devices were returned.-medical records received and evaluation: a review of the follow up notes does not confirm the event of instability.The provided information was reviewed by a consulting clinician and there is no evidence for the causation of instability, further documentation such as operative reports and x-rays would be required to confirm a root cause.-device history review: the device was manufactured and accepted into final stock with no reported discrepancies.-complaint history review: there have been no other events for the lot referenced.Conclusions: revision surgery took place due to mid range instability.The exact cause of the instability could not be determined due to insufficient provision of information.Further information such as: return of reported devices; x-rays, operative reports, and implant retrieval are needed to complete the investigation to determine the root cause of the instability.A capa trend analysis was conducted for the reported failure mode and concluded instability may result from other factors not necessarily related to the device.No further investigation is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Surgeon revised patient with mid range instability and sheered off patella.Left knee.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla navedo
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5197289
MDR Text Key30238491
Report Number0002249697-2015-03592
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2012
Device Catalogue Number5532-G-309
Device Lot Number8YKMHD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/23/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight80
-
-