The explanted generator and lead were received on 01/29/2016.The reported "fracture of lead" allegation was not verified.Scanning electron microscopy images of the connector pin verified that pitting or electro-etching conditions have occurred on the connector pin.An energy dispersive spectrometry (eds) analysis performed on a sample of the reddish-brown deposits observed on the pin showed the presence of some elements present in stainless steel 316.The higher percentage of iron in this sample suggests oxidation may have occurred.An abraded opening of the outer tubing was observed in one location.Since a portion of the lead (including the electrode array) was not returned for analysis, an evaluation and resulting commentary could not be made on that portion of the lead.Other than the above mentioned observations and typical wear and explant related observations, no other anomalies were identified in the returned lead portions.The reported allegation of "high impedance" was not duplicated in the pa lab.Various electrical loads were attached to the pulse generator and results of diagnostic tests demonstrate that accurate resistance measurements were obtained in all instances.In the pa lab, the device output signal was monitored for more than 24-hrs, while the generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator's output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.In addition, a comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery, 3.036 volts, shows an ifi=no condition.In addition, signs of discoloration were also observed on the pulse generator case, which is consistent with the adhered remnants of dried body fluids following an explant process.No other surface abnormalities (such as sharp edges, header delamination, open pockets, decomposition, corrosion or voids) were noted on this device.Observations under pa lab bench microscope revealed pitting on the negative connector block and setscrew.Other than the noted visual anomalies, there were no performance or any other type of adverse conditions found with the pulse generator.
|