Model Number MODEL 250
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
 
Event Description
During a periodic review of the programming history database it was found what appears to be a faulted system diagnostics test that resulted in a change of settings.On (b)(6) 2007 the patient was at settings of output=1.5ma/frequency=20hz/pulse width=250usec/on time=30sec/off time=5min/magnet output=1.75ma/magnet pulse width=250usec/magnet on time=30sec for the final interrogation.On (b)(6) 2008 the patient was found at settings of output=0ma/frequency=20hz/pulse width=500usec/on time=30sec/off time=60min/magnet output=0ma/magnet pulse width=250usec/magnet on time=30sec which appears to have been due to a faulted system diagnostics test.It is unknown when this occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5270420
MDR Text Key33288112
Report Number1644487-2015-06656
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received11/17/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age55 YR
-
-