MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090AA4
Device Problem Issue With Displayed Error Message (2967)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
This report is based solely on device analysis.No information to suggest a device related adverse event or product problem was received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: during analysis it was noted that the device powered up to an error and was further noted that it powered up normally when used with a known good micro processing unit board.Analysis also found that the keyboard cable was damaged.The device was to be scrapped.(b)(4).
 
Event Description
The programmer was returned as a trade-in and was subsequently found during manufacturer's analysis to have its keyboard damaged.It was further reported that it was determined during manufacturer's analysis that the device powered up to an error.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5275337
MDR Text Key33366568
Report Number2182208-2015-04014
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090AA4
Device Catalogue Number2090AA4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received09/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-