• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED SENSOR ENLITE; OZO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED SENSOR ENLITE; OZO Back to Search Results
Model Number MMT-7008A
Device Problems Excess Flow or Over-Infusion (1311); Incorrect, Inadequate or Imprecise Result or Readings (1535); Device Displays Incorrect Message (2591)
Patient Problems Hyperglycemia (1905); Overdose (1988)
Event Date 11/18/2015
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
Customer's mother reported via phone call that the sensors incorrectly alarmed high and low blood glucose levels.Sensor alarmed a 309 mg/dl blood glucose and the insulin pump delivered a correction insulin and caused the customer's blood glucose to drop to 72 mg/dl.Customer's mother reported the sensor triggered the threshold suspend when the customer's blood glucose was 136 mg/d, 122 mg/d, 140 mg/dl.Sensor glucose was 302 mg/dl when the customer's blood glucose was 202 mg/dl.After troubleshooting, customer was advised to monitor the issue.Customer will not be returning the sensors for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE
Type of Device
OZO
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5277837
MDR Text Key33478441
Report Number2032227-2015-72636
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
PMA/PMN Number
120010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial
Report Date 11/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberHG0PSZC
Was Device Available for Evaluation? No
Date Manufacturer Received11/19/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age6 YR
Patient Weight24
-
-