• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE COYOTE¿ ES; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE COYOTE¿ ES; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number H74939135304010
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/30/2015
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).
 
Event Description
It was reported that balloon rupture occurred.Vascular access was obtained via the femoral artery utilizing a contralateral approach.The (b)(6) stenosed target lesion was located in the moderately tortuous and severely calcified left common femoral artery.After a 9-40/0.014x175 non-bsc guidewire crossed the lesion, a 3mmx40mmx146cm coyote (tm) es balloon catheter was advanced for dilation.The balloon was inflated twice at 6 atmospheres.However, on the third inflation at 12 atmospheres for 90 seconds, the balloon ruptured.No segment of the balloon was detached inside the patient after it ruptured.The procedure was completed with 4x4 coyote es catheter balloon.No patient complications were reported and the patient's status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE¿ ES
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5306886
MDR Text Key33774123
Report Number2134265-2015-08770
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
PMA/PMN Number
K080982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 11/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2018
Device Model NumberH74939135304010
Device Catalogue Number39135-30401
Device Lot Number18008622
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/21/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
INTRODUCER SHEATH:SHEATHLESS PV/6FRX55CM/ASASHI
-
-