• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION Back to Search Results
Model Number PE074F5
Device Problem Pacing Problem (1439)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/20/2015
Event Type  malfunction  
Manufacturer Narrative
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product a supplemental report will be sent with the investigation results.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.
 
Event Description
It was reported that the catheter was unable to pace during use.The catheter was replaced and the problem was solved.There were no patient complications reported.
 
Manufacturer Narrative
One bipolar pacing catheter with attached monoject 1.3 cc limited volume syringe was returned for evaluation.Continuity testing confirmed full open conditions of the proximal circuit and the distal circuit.A cut down of the catheter body was performed just proximal of the proximal electrode to expose the pacing leadwires.The proximal and distal leadwires were found to be broken around the solder joints.The balloon inflated clear and concentric with 1.3 cc air and remained inflated for 5 minutes without leakage.No visible damage to the catheter body, balloon, windings or returned syringe was observed.Balloon inflation test was performed using returned syringe with 1.3 cc air by holding the balloon under water.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.The customer report of pacing issue was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION
Type of Device
PACING CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
one edwards way
irvine, CA 92614
9497564386
MDR Report Key5308224
MDR Text Key33833927
Report Number2015691-2015-03429
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPE074F5
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/15/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-