Model Number PE074F5 |
Device Problem
Pacing Problem (1439)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/20/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product a supplemental report will be sent with the investigation results.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.
|
|
Event Description
|
It was reported that the catheter was unable to pace during use.The catheter was replaced and the problem was solved.There were no patient complications reported.
|
|
Manufacturer Narrative
|
One bipolar pacing catheter with attached monoject 1.3 cc limited volume syringe was returned for evaluation.Continuity testing confirmed full open conditions of the proximal circuit and the distal circuit.A cut down of the catheter body was performed just proximal of the proximal electrode to expose the pacing leadwires.The proximal and distal leadwires were found to be broken around the solder joints.The balloon inflated clear and concentric with 1.3 cc air and remained inflated for 5 minutes without leakage.No visible damage to the catheter body, balloon, windings or returned syringe was observed.Balloon inflation test was performed using returned syringe with 1.3 cc air by holding the balloon under water.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.The customer report of pacing issue was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.
|
|
Search Alerts/Recalls
|