• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RELIANT STENT GRAFT BALLOON CATHETER -OUS; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RELIANT STENT GRAFT BALLOON CATHETER -OUS; CATHETER, PERCUTANEOUS Back to Search Results
Model Number REL46J
Device Problems Inability to Irrigate (1337); Component or Accessory Incompatibility (2897)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/17/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was returned for analysis.The events were confirmed; there was difficult passing a guidewire and difficulty flushing the reliant balloon.The root cause of the event was most likely related to the physician¿s decision to cut the guidewire after part of it had become contaminated during the procedure and then trying to re-insert the cut end back through the guidewire lumen.The cut end of the guidewire then scraped along the inner lumen causing the material to build up as the guidewire was passed and ultimately led to the occlusion at the distal tip.
 
Event Description
A reliant balloon catheter was selected as an accessory device for the treatment of an abdominal aortic aneurysm.It was reported that the during the procedure, after the guidewire was inserted through the reliant balloon without issue, the guidewire had become dirty.The physician removed the guidewire and cut off a part of the contaminated wire.The physician then tried to insert the cut end of the guidewire back through the reliant balloon; however, had difficulty in insertion into guide wire lumen.The physician then attempted to flush the lumen however, was unsuccessful.No clinical sequelae were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANT STENT GRAFT BALLOON CATHETER -OUS
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC CARDIOVASCULAR
3576 unocal place
santa rosa CA 95403
Manufacturer Contact
eric elliott
3576 unocal place
santa rosa, CA 95403
7075912586
MDR Report Key5310848
MDR Text Key34612210
Report Number2953200-2015-02385
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/14/2017
Device Model NumberREL46J
Device Catalogue NumberREL46J
Device Lot Number0007731195
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/24/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-