• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE STERLING¿; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE STERLING¿; CATHETER, PERCUTANEOUS Back to Search Results
Model Number H74939031604010
Device Problem Material Rupture (1546)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/03/2015
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr: the complaint device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and moderately calcified superior femoral artery.A 6.0mmx40mmx135cm sterling b alloon catheter was advanced to the lesion.The balloon was inflated for 5 seconds however it ruptured at 6 atmospheres on the first inflation.The ruptured balloon was completely removed from the patient.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING¿
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5322218
MDR Text Key34920653
Report Number2134265-2015-08868
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
PMA/PMN Number
K053118
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 12/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/02/2016
Device Model NumberH74939031604010
Device Catalogue Number39031-60401
Device Lot Number16580202
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/03/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/04/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial