Model Number MERGE CARDIO 9.0.1 |
Device Problem
Patient Data Problem (3197)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/11/2012 |
Event Type
malfunction
|
Manufacturer Narrative
|
Upon retrospective review, this issue was determined to be reportable as an mdr.
|
|
Event Description
|
On (b)(6) 2012 a merge cardio customer reported reports done under wrong patient.Report from 2nd patient is also showing up as a 2nd report thread for patient #1.The pdf generation functionality was not designed to be exclusive per patient.If two or more users were reporting on different patients and requested the generation of a pdf concurrent, the system would incorrectly attach the first pdf report file generated to the other patient(s).
|
|
Manufacturer Narrative
|
The correction report (b)(4) that was referenced in the initial mdr was cancelled by fda.
|
|