Model Number ENB-375-20-135 |
Device Problem
Burst Container or Vessel (1074)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/03/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
A review of the manufacture records for this device is ongoing.Once the evaluation is complete a supplemental report will be submitted.(b)(4).
|
|
Event Description
|
During the procedure the enteer balloon burst.The balloons on the enteer burst when it was inserted a second time and the wire would not re-enter the lumen.The procedure was aborted.The lesion being treated was 100% cto in a 7mm diameter artery.The lesion length was 250mm.The balloon was at 8atm when it burst (it is indicated at 3-4 atms).No injury reported.The physician is planning to bring the patient back another day.
|
|
Manufacturer Narrative
|
Evaluation of the returned enteer catheter was completed january 14, 2016.A review of the manufacture records could not be conducted because the lot number provided was not for the enteer device.The enteer catheter was received without the enteer guidewire.The enteer catheter was inspected and found blood within the balloon and catheter shaft of the device.The balloon was inspected under microscope and both guidewire exit ports were undamaged and no signs of obstruction were observed.Both marker bands were intact and accounted for.A test syringe was used to inflate the balloon.A leak was identified at the distal end of the balloon.After flushing the balloon a longitudinal tear was identified.The customer¿s report of a balloon burst has been confirmed.A visual inspection of the balloon identified a longitudinal tear at the distal end of the balloon after being identified during functional testing.The root cause could not be identified.
|
|
Search Alerts/Recalls
|