MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; OZO Back to Search Results
Model Number MMT-751NAH
Device Problems Crack (1135); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/24/2015
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.(b)(4).
 
Event Description
The customer reported via phone call of a crack on the insulin pump.The customer's blood glucose reading was unknown.The customer stated that she noticed the crack about 3 weeks ago.The customer stated that the crack is near the battery compartment.The customer stated that she had issues with the failed battery test.The customer was advised to disconnect from the pump.The customer was advised to discontinue use of the device and revert to a backup plan per health care provider's instructions.The customer will be sent a replacement pump.
 
Manufacturer Narrative
The insulin pump was received with corroded battery tube however, was monitored and no unexpected failed battery test alarms occurred during our testing.The insulin pump received with normal operating currents.The insulin pump had cracked reservoir tube lip, broken battery tube threads, minor scratched lcd window and scratched reservoir tube window.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP
Type of Device
OZO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5339562
MDR Text Key35248865
Report Number2032227-2015-80332
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
PMA/PMN Number
P120010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 12/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751NAH
Device Catalogue NumberMMT-751NAH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received12/15/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age53 YR
Patient Weight79
-
-