• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE FLORIDA OPERATIONS PERFECTO2 V WITH SENSOR 9153650799; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE FLORIDA OPERATIONS PERFECTO2 V WITH SENSOR 9153650799; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number IRC5PO2V
Device Problems Device Alarm System (1012); Leak/Splash (1354); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
A return was issued and the product is awaiting receipt and/or evaluation. a follow up will be filed if/when any additional information is provided.
 
Event Description
The unit had no alarms, compressor was grinding, and there was a sieve leak.
 
Manufacturer Narrative
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the compressor won't run/start, which confirmed the original complaint issue for the compressor.The complaint issue for unit has no alarms and sieve leak were not confirmed.However, the underlying cause could not be determined.
 
Event Description
The unit had no alarms, compressor was grinding, and there was a sieve leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFECTO2 V WITH SENSOR 9153650799
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5346286
MDR Text Key35546904
Report Number1031452-2016-00023
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Replace
Type of Report Initial,Followup
Report Date 01/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberIRC5PO2V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received01/25/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-