Model Number IRC5PO2V |
Device Problems
Device Alarm System (1012); Leak/Splash (1354); Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
A return was issued and the product is awaiting receipt and/or evaluation. a follow up will be filed if/when any additional information is provided.
|
|
Event Description
|
The unit had no alarms, compressor was grinding, and there was a sieve leak.
|
|
Manufacturer Narrative
|
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the compressor won't run/start, which confirmed the original complaint issue for the compressor.The complaint issue for unit has no alarms and sieve leak were not confirmed.However, the underlying cause could not be determined.
|
|
Event Description
|
The unit had no alarms, compressor was grinding, and there was a sieve leak.
|
|