The catheter and guidewire were returned for analysis.The stylet was found protruding from the catheter hub.No damages were observed on the catheter body.The stylet was removed from within the catheter without issue.No damages were found with the stylet.The catheter was flushed and found patent.The catheter did not have leaks, ruptures, or punctures.The guidewire was examined and the outer coils were found stretched, with the core wire intact.However, there was damaged and missing coating from the proximal segment.The guidewire tip was found bent at the distal segment.Due to the damaged condition of the guidewire, it could not be used for further testing.An in-house guidewire was able to be navigated through the catheter lumen without issue.No other anomalies were observed.Based on the reported information and the device analysis, the customer¿s report of ¿guidewire puncture¿ could not be confirmed.No damages or issues were found with the returned catheter.The catheter performed as intended with an in-house guidewire.However, the returned guidewire was found to be damaged.The cause for the damage to the guidewire and the reported experience could not be determined.Mdr related to this event: 2029214-2016-00018 2029214-2016-00019.
|