U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS MONITORING SYSTEM DEXCOM SHARE SYSTEM; MDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS MONITORING SYSTEM DEXCOM SHARE SYSTEM; MDS Back to Search Results
Model Number MT22495
Device Problems Self-Activation or Keying (1557); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/16/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Patient contacted dexcom on (b)(6) 2015 to report that the receiver displayed initializing screen without manual restart on (b)(6) 2015.At the time of contact, no injury or medical intervention was reported.The device has been received for evaluation.A follow-up report will be submitted once the evaluation is complete.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.A screen error alarm was observed during a review of the downloaded data log.The reported event of a hardware failure was confirmed.A root cause could not be determined.The dexcom g4 platinum continuous glucose monitoring system rev.11, under section 13.8.2 receiver error code notes the following: this screen shows an error code that means the receiver may not be working properly.Write down the error code and contact dexcom technical support.Continue to check your blood glucose value using your blood glucose meter.No alert sound or vibration will warn you that you are no longer getting sensor glucose readings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS MONITORING SYSTEM DEXCOM SHARE SYSTEM
Type of Device
MDS
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key5361365
MDR Text Key35729025
Report Number3004753838-2016-30284
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000156
UDI-Public(01)00386270000156(241)MT22495(10)5206726(17)NI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495
Device Catalogue NumberSTK-DR-001
Device Lot Number5206726
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/11/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received01/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
Patient Weight61
-
-