MAUDE Adverse Event Report: ABBOTT LABORATORIES ARCHITECT CA 19-9XR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES ARCHITECT CA 19-9XR Back to Search Results
Catalog Number 02K91-28
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process. A follow up report will be submitted when the evaluation is complete. (b)(4). An evaluation is in process.
 
Event Description
The customer observed falsely elevated ca19-9 results on the architect i2000sr analzyer. The following data was provided for a patient diagnosed with pancreatic cancer in 2011: (b)(6) previous results of less than 10, repeats 128 (obtained on instrument (b)(4)), 113 (obtained on instrument (b)(4)). There was no impact to patient management reported.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and accuracy testing. No adverse trend was identified for the customer's issue. Labeling was reviewed and found to be adequate. The product was not available for return. An internal panel was tested with retained kits of the likely cause reagent lot and accuracy testing met all specifications. Based on all available information and abbott diagnostics' complaint investigation, the assay performed as intended and no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCHITECT CA 19-9XR
Type of DeviceCA 19-9
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 3500
Manufacturer (Section G)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 3500
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key5361532
MDR Text Key35991012
Report Number1415939-2016-00019
Device Sequence Number1
Product Code NIG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K052000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/04/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/12/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date07/15/2016
Device Catalogue Number02K91-28
Device Lot Number50371M500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-