Product analysis for the returned generator demonstrated that accurate resistance measurements were obtained in all instance using various electrical loads.A comprehensive automated electrical evaluation showed that the vns generator performed according to functional specifications.The final electrical test showed an ifi = no condition.There were no performance or any other type of adverse conditions found with the vns generator.Product analysis for the returned lead found abraded an opening on the inner silicone tube, in one locations, and abraded openings on the outer silicone tube.The abraded opening on the outer tubing and the cut ends made during the explant process most likely provided the leakage path for the dried remnants of what appeared to have been body fluids found inside the outer tubing.What appeared to be white deposits were observed in various locations.Energy dispersion spectroscopy was performed on the deposits observed on the connector boot and identified the deposit as containing silicon and sodium.With the exception of the abraded openings, the condition of the returned lead portions is consistent with those that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide the evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed during visual analysis and no discontinuities were identified.Based on these findings in the product analysis lab, there was no evidence to support the low impedance, short circuit, or fractured lead allegations.It should be noted that a portion of the lead assembly, including the electrodes, was not returned for analysis; therefore, a complete evaluation could not be performed on the entire lead product.
|
It was also reported the patient had an increase in seizures, above pre-vns baseline levels, which began after the lead fracture occurred.The increased seizures continued once the new devices were implanted and was reported, and will continue to be investigated, in mfr.Report 1644487-2016-01782.
|