MAUDE Adverse Event Report: INVAMEX TRANSPORT 20 IN X 16 IN 9153637574; WHEELCHAIR, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVAMEX TRANSPORT 20 IN X 16 IN 9153637574; WHEELCHAIR, MECHANICAL Back to Search Results
Model Number T422RFAP
Device Problems Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available a supplemental record will be filed.
 
Event Description
The dealer stated the rear wheel (not sure which side) was hitting the frame.They tried to change the rear wheel and it continued to hit the frame.He stated there is an issue with the frame of the chair.
 
Manufacturer Narrative
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the left rear arm gusset doesn't line up correctly not allowing enough clearance between left rear wheel and left arm, which confirmed the original complaint issue.However, the underlying cause could not be determined.
 
Event Description
The dealer stated the rear wheel (not sure which side) was hitting the frame.They tried to change the rear wheel and it continued to hit the frame.He stated there is an issue with the frame of the chair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSPORT 20 IN X 16 IN 9153637574
Type of Device
WHEELCHAIR, MECHANICAL
Manufacturer (Section D)
INVAMEX
parque industrial manimex
reynosa 88780
MX  88780
Manufacturer (Section G)
INVAMEX
parque industrial manimex
reynosa 88780
MX   88780
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5383427
MDR Text Key36536214
Report Number9616091-2016-00086
Device Sequence Number1
Product Code IOR
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Replace
Type of Report Initial,Followup
Report Date 01/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberT422RFAP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received01/21/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-