Model Number FA-71500-25 |
Device Problem
Positioning Failure (1158)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/18/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has not been returned for evaluation; however, return is anticipated.Without return of the device, no definitive conclusions can be drawn regarding the clinical observation.Should the device be returned or additional information become available, a supplemental report will be submitted.
|
|
Event Description
|
Medtronic received information that during treatment of aneurysm located in the right cavernous sinus, ophthalmic artery, the device could not be released from the capture coil.It was reported the physician rotated the pushwire more than ten times in effort to release the device but was unsuccessful.The device was then retracted and another device was used to complete the surgery successfully.No patient injury was reported.
|
|
Manufacturer Narrative
|
The pipeline braid was returned for evaluation without the pushwire.As received the distal and proximal ends of the pipeline braid could not be determined.The pipeline braid was observed to be fully open with both ends having slight fraying.No other anomalies were observed.Based on the analysis findings the clinical observation could not be confirmed.We are unable to definitively determine the cause for the reported experience.However, it is possible that the rotating of the pushwire more than the recommended ten times, and the damaged braid may have contributed to reported experience.However, the cause for the damage could not be determined.As per the instructions for use (ifu): ¿never rotate the delivery wire more than 10 full turns.If ped does not open after 10 turns, remove the entire system (micro catheter and ped delivery system together).¿ all products are 100% inspected for damage and irregularities during manufacture.
|
|
Search Alerts/Recalls
|